# Class II recall Z-0119-2027: Intellijoint Surgical, Inc.

**Predicate Speed & CAPA Advisory Opportunity Score:** 62/100

| Field | Value |
|---|---|
| Record | Device recall |
| Recall number | Z-0119-2027 |
| Recalling firm | Intellijoint Surgical, Inc. |
| Classification | Class II |
| Status | Ongoing |
| Voluntary / mandated | Voluntary: Firm initiated |
| Specialty | Other |
| Product | Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249 |
| Reason | A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk. |
| Quantity | 8 |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain. |
| Recall initiated | 2026-08-28 |
| Reported | 2026-10-14 |
| Firm location | Kitchener, N/A, Canada |

## Why this score
- Class II recall: usually triggers CAPA and quality-system remediation work.
- Recall is still ongoing.
- Posted 22 days ago.

Source: U.S. FDA openFDA (510(k) clearances and device enforcement reports) (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0119-2027%22). Last refreshed: 2026-11-04T21:30:53.898Z.