# Class II recall Z-0151-2027: Baxter Healthcare Corporation

**Predicate Speed & CAPA Advisory Opportunity Score:** 62/100

| Field | Value |
|---|---|
| Record | Device recall |
| Recall number | Z-0151-2027 |
| Recalling firm | Baxter Healthcare Corporation |
| Classification | Class II |
| Status | Ongoing |
| Voluntary / mandated | Voluntary: Firm initiated |
| Specialty | Other |
| Product | Likorall Overhead Lifts    Product Code: 3121001  Product Description: LIKORALL 200    Product Code: 3122005  Product Description: LIKORALL 242 ES, NATURAL    Product Code: 3122005NA  Product Description: LIKORALL 242 ES, NATURAL    Product Code: 3122006	  Product Description: LIKORALL 242 ES, WHITE	    Product Code: 3122006NA  Product Description: LIKORALL 242 ES, WHITE	    Product Code: 3122007  Product Description: LIKORALL 242 ES R2R, NATURAL    Product Code: 3122007NA	  Product Description: LIKORALL 242 ES R2R, NATURAL	    Product Code: 3122008  Product Description: LIKORALL 242 ES R2R, W |
| Reason | Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue. |
| Distribution | Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi,  |
| Recall initiated | 2026-09-14 |
| Reported | 2026-10-21 |
| Firm location | Deerfield, IL, United States |

## Why this score
- Class II recall: usually triggers CAPA and quality-system remediation work.
- Recall is still ongoing.
- Posted 15 days ago.