# Class II recall Z-0203-2027: Medtronic MiniMed, Inc.

**Predicate Speed & CAPA Advisory Opportunity Score:** 68/100

| Field | Value |
|---|---|
| Record | Device recall |
| Recall number | Z-0203-2027 |
| Recalling firm | Medtronic MiniMed, Inc. |
| Classification | Class II |
| Status | Ongoing |
| Voluntary / mandated | Voluntary: Firm initiated |
| Specialty | Other |
| Product | CareLink Clinic, REF: MMT-7350 |
| Reason | CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypo |
| Quantity | 1 |
| Distribution | US Nationwide distribution in the state of KS. |
| Recall initiated | 2026-09-11 |
| Reported | 2026-10-28 |
| Firm location | Northridge, CA, United States |

## Why this score
- Class II recall: usually triggers CAPA and quality-system remediation work.
- Recall is still ongoing.
- Posted 8 days ago.