# Class II recall Z-0236-2027: Boston Scientific Cork, Ltd.

**Predicate Speed & CAPA Advisory Opportunity Score:** 68/100

| Field | Value |
|---|---|
| Record | Device recall |
| Recall number | Z-0236-2027 |
| Recalling firm | Boston Scientific Cork, Ltd. |
| Classification | Class II |
| Status | Ongoing |
| Voluntary / mandated | Voluntary: Firm initiated |
| Specialty | Other |
| Product | Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/155;  Material Number: M001195330 |
| Reason | The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. |
| Quantity | 206 |
| Distribution | Worldwide distribution   US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Jap |
| Recall initiated | 2026-09-16 |
| Reported | 2026-10-28 |
| Firm location | Cork, Ireland |