# Class II recall Z-3231-2026: Biomerieux Inc

**Predicate Speed & CAPA Advisory Opportunity Score:** 62/100

| Field | Value |
|---|---|
| Record | Device recall |
| Recall number | Z-3231-2026 |
| Recalling firm | Biomerieux Inc |
| Classification | Class II |
| Status | Ongoing |
| Voluntary / mandated | Voluntary: Firm initiated |
| Specialty | Other |
| Product | VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811 |
| Reason | The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health. |
| Quantity | 97680 units |
| Distribution | International distribution to the countries of Austria, Germany, Luxembourg. |
| Recall initiated | 2026-07-24 |
| Reported | 2026-09-30 |
| Firm location | Hazelwood, MO, United States |